An automatic analyzing apparatus for effecting chemical analyses for various sample liquids such as blood, urine, and the like, comprising a sample delivery pump for metering a sample liquid into a reaction cuvette, a reagent delivery pump for delivering to the reaction cuvette a given amount of a given reagent selected from a plurality of reagents contained in a reagent cassette, to form a test liquid, a feed mechanism for successively supplying reaction cuvettes along a circular reaction line, a plurality of photometering sections arranged along the reaction line for effecting a plurality of photometric and/or nephelometric and/or fluorometric measurements for each test liquid at different time instances to produce a plurality of photometric results, and circuitry for receiving the photometric results and selecting therefrom given quantitative analytical data of a given test item.
Centrifugal analyzers are also discrete but test liquids are processed in parallel to one another. All samples in process are in the same stage of analysis at the same time. In operation, samples and reagent are pre-measured and pre-loaded into appropriate compartments arranged about the circumference of a rotor disc, whereupon it is placed on a centrifuge and rotated at a high speed past a photometer device. Centrifugal force mixes all samples with reagent at the same time and hence each of the test liquids is in the same stage of analysis at any given point in time.
The majority of analyzers, regardless of the above mentioned categories, are capable of performing more than one type of test item. There are three broad categories of methods for providing for multi-test capability.
What shall hereinafter be referred to as Random Access Testers currently require individual test packs which are pre-packaged with the appropriate reagents required to perform one test of a given test type. These test packs are loaded into the instrument system according to the analyst's needs, charged with a sample liquid, and processed in a discrete manner. Random access testers offer great convenience and flexibility but currently available embodiments have low productivities when compared with other means of providing multi-test item capability. In addition, the requirement for pre-packaged test packs makes operating costs much higher than the alternate methods.
Another means of performing a plurality of tests on each of a plurality of samples is sequentially by test-item batch. All samples are analyzed sequentially or centrifugally for a given test item. When all samples have been analyzed for a given test item, the system is changed over, or somehow modified, to perform a different test item and all appropriate samples are re-treated. When all samples have been processed for the required test items, the results of each sample's test items must be collated to allow including all of a given samples analytical results on a single report form for return to a physician, etc. Such systems are usually referred to as `single channel` systems. Single channel systems are usually considered most appropriate for treating a batch or plurality of samples, as the effort required to change-over from one test item to another is generally neither convenient nor cost-effective to treat one sample for a plurality of test items. Additionally, at any given moment in time, only one test item is available for immediate use.
Simultaneous analyzers have a plurality of analytical channels which enable a plurality of test items to be performed simultaneously on each sample. Such systems are commonly referred to as `multi channel` analyzers. Multi-channel analyzers do make more than one test item available at any given point in time, do eliminate the data collating task required of single channel analyzers and in general, do have higher productivities than single-channel analyzers by virtue of the fact that they are constructed as a plurality of single-channel analyzers combined into one device. This last feature is a drawback in that it makes the analyzer system complicated in construction, large in size, and generally, much higher in cost than single-channel discrete, continuous flow or centrifugal analyzers.
In the known analytical systems of the non-centrifugal type, photometric quantitation is carried out after some time period from the initiation of the test reaction, i.e. when the test liquid has traveled along the processing line by some given fixed distance. Therefore, the reaction time is fixed as a function of the length or circumference of the processing line, which may or may not be optimal with respect to a given test item and/or sample.
Additionally, sequential testers have only one photometer position per channel, severely limiting the amount of photometric data which can be made available. No photometric data can be made available until a test liquid reaches the photometer station, typically, 8-10 (often 30) minutes from the time of mixing of sample with reagent. Once a test liquid reaches a photometer station, the amount of time which is devoted to photometric measurement essentially limits the speed of analysis of a given sequential tester, i.e. if 60 seconds is devoted to photometric quantitation, then the processing rate is limited to 60 tests per hour. This feature forces a trade-off between processing rate and photometric quantitation time especially for `kinetic` test (ex. enzyme rate tests) which require photometric measurement over long periods of time in order to provide for best accuracy and precision of analysis.
SUMMARY OF THE INVENTION
The present invention has as its object to provide for an automatic analyzing apparatus which is so constructed that the above drawbacks can be avoided while insuring consistently reliable results.
According to the invention, an apparatus for effecting automatic analysis comprises means for successively feeding reaction vessels, each containing a respective test liquid to be analyzed, along a given reaction line;
means for delivering (a) given amount(s) of (a) given reagent(s), corresponding to a test item to be measured, into a reaction vessel on the reaction line to form a test liquid;
a plurality of photometering means arranged at different measuring positions distributed along the reaction line for effecting a plurality of photometric measurements for a respective test liquid in a vessel at different time instances;
means for receiving results of said plurality of photometric measurements and selecting therefrom given quantitative analytical data of a given test item for the test liquid in a reaction vessel; and
means for discharging the test liquid out of the reaction vessel after the quantitative analysis for the given test item has been performed.
BRIEF DESCRIPTION OF THE DRAWINGS
FIG. 1 is a schematic view illustrating a principal construction of an automatic analyzing apparatus according to the invention;
FIG. 2 is a graph showing typical reaction state of a test liquid;
FIGS. 3 and 4 are perspective views illustrating an embodiment of the automatic analyzing apparatus of the invention;
FIG. 5 is a schematic plan view showing an arrangement of various portions of the apparatus shown in FIGS. 3 and 4;
FIG. 6 is a plan view showing a photometering section of the apparatus of FIG. 5;
FIG. 7 is a schematic cross-sectional view showing the photometering section;
FIG. 8 is a plan view showing a rotary filter unit illustrated in FIGS. 6 and 7;
FIGS. 9A and 9B are graphs showing reaction curves;
FIG. 10 is a chart for explaining an operation of the apparatus according to the invention;
FIG. 11 is a perspective view showing an embodiment of a cuvette for use in the apparatus according to the invention;
FIGS. 12A and 12B are side views illustrating a manner of holding the cuvette of FIG. 11;
FIG. 13 is a plan view showing an embodiment of a reagent cassette;
FIG. 14 is a perspective view illustrating the reagent cassette;
FIG. 15 is a block diagram showing a manner of controlling a reagent feed mechanism for minimizing a total travelling distance of the reagent cassette;
FIG. 16 is a schematic view illustrating an embodiment of the cassette holder comprising separate refrigerator and room temperature portion;
FIG. 17 is a perspective view showing an embodiment of the refrigerator of FIG. 16;
FIG. 18 is a schematic view explaining a delivery operation of the reagent delivery mechanism shown in FIG. 5;
FIG. 19 is a schematic view showing an embodiment of the reagent delivery mechanism;
FIGS. 20A and 20B are a perspective view and a graph, respectively, showing an embodiment of a liquid level detector of the reagent delivery mechanism and a relation between an amount of sucked liquid and a detection output, respectively;
FIGS. 21A and 21B are a cross-sectional view and a graph showing another embodiment of the liquid level detector;
FIGS. 22A and 22B show still another embodiment of the liquid level detector;
FIG. 23 is a perspective view illustrating an embodiment of a liquid level detector for a reagent in a reagent bottle;
FIG. 24 is a perspective view showing another embodiment of the reagent level detector;
FIG. 25 is a schematic view showing an embodiment of a probe washing device;
FIG. 26 is a perspective view depicting another embodiment of the washing device;
FIGS. 27A and 27B are schematic views for explaining the operation of the washing device shown in FIG. 26;
FIG. 28 is a block diagram showing a manner of connecting the automatic analyzing apparatus according to the invention to a computer installed at a hospital;
FIG. 29 is a block diagram illustrating a manner of coupling the apparatus according to the invention with a back-up computer through a communication line;
FIG. 30 is a block diagram showing a manner of controlling or operating a plurality of the analyzing apparatuses according to the invention by means of a single controlling unit;
FIG. 31 is a flow chart showing an embodiment of a patient data system using the apparatus according to the invention;
FIG. 32 is a flow chart showing another embodiment of the patient data system;
FIG. 33 is a plan view showing a format of a patient card for use in the patient data system;
FIG. 34 is a schematic view showing an embodiment of a cuvette and liquid discharging mechanism;
FIG. 35 is a schematic view showing another embodiment of the discharging device;
FIG. 36 is a schematic view showing still another embodiment of the discharging device;
FIG. 37 is a schematic view showing another embodiment of the apparatus according to the invention;
FIG. 38 is a schematic view showing an embodiment of an ion concentration measuring device which may be installed in the apparatus according to the invention;
FIG. 39 is a schematic view showing another embodiment of the ion concentration measuring device;
FIG. 40 is a block diagram showing an embodiment of a signal processing circuit of the ion concentration measuring devices shown in FIGS. 38 and 39;
FIG. 41 is a cross section showing schematically an embodiment of a photometric section of the apparatus according to the invention, which can effect colorimetric, nephelometric and fluorometric analyses;
FIG. 42 is a cross section showing schematically another embodiment of the photometric section;
FIG. 43 is a perspective view showing an embodiment of a shutter mechanism shown in FIG. 42;
FIG. 44 is a side view illustrating an embodiment of the cuvette;
FIGS. 45A and 45B are schematic views illustrating an embodiment of the photometric section in which the transmitted, scattered and fluorescent lights are received by a single light receiving element;
FIG. 46 shows another embodiment of the photometric section;
FIG. 47 illustrates still another embodiment of the photometric section; and
FIGS. 48A and 48B are schematic views illustrating another embodiment of the photometric section in which the scattered, transmitted and fluorescent lights can be received by a single element by using cuvettes having different configurations.
DESCRIPTION OF THE PREFERRED EMBODIMENTS
FIG. 1 is a schematic view illustrating a constructional principle of the automatic analyzing apparatus according to the invention. This apparatus can be classified as a discrete system adopting a batch process and belongs to a sequential multi system in which analyses for a plurality of test items can be effected continuously in succession. Sample vessels 1 are supported on a sample feed mechanism 2 and are intermittently fed in a direction shown by an arrow A. A given amount of sample liquid, contained in the successive sample vessels 1 are aspirated by a sample delivery mechanism 3 at a given position in accordance with test items to be analyzed and the given amount of sample liquid is supplied into cuvettes 4 together with a diluent 5 as required. The cuvettes 4 are supported by a cuvette feed mechanism 6 and are intermittently fed along a reaction line B in a direction shown by an arrow B at a predetermined period, such as six seconds per step. New cuvettes 4 are successively supplied to the feed mechanism 6 from a cuvette-delivery mechanism 7. The cuvette 4 having the sample liquid delivered therein is advanced by several steps and arrives at a given position at which point a reagent, dependent on the test item to be measured, is delivered in the cuvette 4 together with a diluent 9 by means of a reagent-delivery mechanism 8. Reagents to be used for measurement are contained in reagent bottles 10.sub.1 -10.sub.n which are supported on a reagent feed mechanism 11 movable in a reciprocal manner as shown by a double headed arrow C. A given reagent can be drawn by the delivery mechanism 8 from the bottle which is positioned at the given delivering position. The sample liquid and reagent can be sufficiently mixed by jetting the reagent into the cuvette 4 together with the diluent at a suitable flow rate. The cuvette 4 having had reagent and sample delivered thereto travels along the reaction line B. The test liquid in the cuvette is measured by photometers 12 to 15 each comprising a light source L and a light-receiving element S provided at positions separated from each other by distances equal to multiples of a traveling step of the cuvette. In this manner the reaction state of the test liquid in the cuvette 4 can be monitored as it progresses along the reaction line.
Particularly in a measurement of enzymatic reactions, it is very important to monitor the reaction over some extended period of time. That is to say, in the measurement of enzymatic reactions, it is impossible to obtain an accurate result unless a measurement is effected during the linear portion of an absorbance level-to-time characteristic curve. In FIG. 2, a typical reaction curve is shown and an absorption (O.D.) is plotted on the ordinate and time (t) measured from the addition of reagent, is plotted on the abscissa. In FIG. 2, a left-hand zone (a) represents the lag phase of reaction due to heating time of test liquid, mixing, etc., and a zone (b) denotes the linear phase in which the reaction rate measurement, i.e. kinetic reaction measurement, can be effected positively and accurately. Further a zone (c) represents an end point phase in which the reagent substrate or other given components in the test liquid have been exhausted. Measurement in the end point zone (c) results in erroneously low values when performing kinetic assays. The period of the linear phase (b) may be suitably changed by adjusting the substrate concentration, etc. and total volume of test liquid. This adjustment is effected in such a manner that the end of the lag phase (a) can be detected by the photometers 12 to 15 (see FIG. 1) for almost all test liquids even if the test liquids have fast or slow reaction rate. Preferably the substrate concentration conditions, and total volume of test liquid are so adjusted that the variation in absorption can be observed after twelve seconds (corresponding to the position of photometer 12) from the mixing of reagent and sample for the test liquid having the slowest reaction rate and the linear phase (b) will last for one or two minutes or more for the normal test liquids. By such a measure, the lag phase of successively fed test liquids can be monitored in a substantially completed state by the photometers 12 to 14. It should be noted that the photometers 12 to 15 can monitor the linear phase (b) as well as the lag phase (a). That is to say, when the end of the lag phase is detected for a test liquid by one of the photometers 12 to 14, the measurement is effected for the relevant test liquid during the linear phase by means of a photometer which is situated beyond the above mentioned photometer on the reaction line. After the measurement, the test liquid is discharged together with the cuvette 4.
The above mentioned sample feed mechanism 2, sample delivery mechanism 3, cuvette feed mechanism 6, reagent delivery mechanism 8, reagent feed mechanism 11 and the photometering sections can be controlled by a control device 16 including a computer on the basis of patient information introduced by an operator.
As described above, according to one aspect of the invention, the lag phase and linear phase are monitored at a number of positions on the reaction line to obtain a number of photometric data and then useful data are selectively derived from these data. By this measure it is possible to obtain the analytical data of high accuracy and reliability and thus a useful automatic analyzing apparatus having excellent and unique abilities can be realized.
Now, embodiments of the apparatus according to the invention will be explained.
FIGS. 3 and 4 are perspective views illustrating an outer appearance of the automatic analyzing apparatus according to the invention. A main body 25 includes a cover 26 hinged at the rear to provide access to internal components. In the cover 26 are formed openings 27A and 27B for dissipating heat produced by light sources of photoelectric colorimeters. A front plate 28 is secured to the main body 25 in such a manner that the front plate can be opened to provide access. A cuvette container 29 for storing waste cuvettes and a waste liquid container 30 for storing waste liquid are detachably secured to the front plate 28. A right-hand side plate 31 is hinged to the main body 25 at the bottom side and a cassette holder 32 for supporting a detachable reagent cassette for holding various reagent bottles necessary for given analyses is provided on the side plate 31. A portion for fitting the cassette holder 32 defined by the right hand side plate 31 forms a refrigerator 33.
A sample liquid feed mechanism 34 is affixed to the main body 25 at its front portion. This mechanism comprises a rotating gear-like turntable which can be detachably installed in the apparatus when the cover 26 is opened. As shown in FIG. 4, a chain may be engaged with the turntable so as to feed the sample cups held by the chain. This chain may be selectively used, depending upon the number of test bodies to be analyzed.
FIG. 5 is a schematic view illustrating an arrangement of various portions of the apparatus with the top cover 26 removed. The sample cups are fed successively by the feed mechanism 34 to a given aspiration position. The cuvettes are fed one by one by a cuvette supply mechanism 35 through a position near the sample aspiration position. A given amount of sample from the sample cup is supplied to the cuvette by a pump 36. While the cuvette is fed to a photometric position by a cuvette feed mechanism 37, a given amount of a suitable reagent is supplied to the cuvette from a reagent bottle 38 in the reagent cassette 32 by means of a reagent dispenser 39. A plurality of reagent bottles 38 are arranged in the cassette 32 along an endless path and any desired bottle 38 can be indexed at a position for aspirating the reagent therein by the dispenser 39. As will be explained later in detail, an ion sensor 40 is arranged along the cuvette feed mechanism 37 to measure concentrations of ions in the test liquid. At the end of the cuvette feed mechanism 37 a distributing mechanism 41 is arranged for continuously delivering successive cuvettes into right and left photometering sections 42A and 42b alternately. These photometering sections are communicated with the openings 27A and 27B, respectively formed in the cover 26. After the measurement in the sections 42A, 42B, the cuvette and its container are discharged at stations 43A and 43B.
When two photometering assembles 42 are provided even if the cuvettes are successively fed every six seconds by means of the feed mechanism 37, each test liquid can be measured for twelve seconds at each photometer position by either one of the photometering sections and thus, the time available for measurement can be increased. Further, even if one of the photometering assemblies becomes inoperative, the analyzing operation can be carried out.
Next, a detailed construction of the photometering section will be explained. As illustrated in FIGS. 6 and 7 the photometering section 42 comprises a disc-shaped turntable 44 surrounding a chimney 27. A plurality of cuvettes are arranged along the periphery of the turntable 44. These cuvettes can be indexed past a number of photometering positions. The cuvette 45 is made at least partially of transparent material. A single light source 46 is arranged in the chimney 27 and a number of apertures 47 are formed in a cylindrical body defining the chimney 27 at positions corresponding to a number of photometering positions. These apertures are situated at the same vertical level as the light source 46. Around the cylinder forming the chimney 27 is rotatably arranged a drum 49 having formed a pair of slits 48 therein at the same level as the apertures 47. The The drum 49 is rotated by a motor 50 at a high speed. There are further arranged a number of optical fibers 51 each having one end secured at a respective photometering position so as to receive a light emitted from the light source 46 through the aperture 47 and the slit 48 and transmitted through the cuvette 45. The other ends of these respective fibers are collected at one or two positions and are faced to photo detectors 52 comprising a photomultiplier tube or similar device. Between the collected ends of the fibers 51 and the photo detector 52 is arranged a rotary filter unit 53. As shown in FIG. 8, the rotary filter unit 53 comprises a plurality of filters .lambda..sub.1 to .lambda..sub.10 having different transmitting wavelengths and is rotated by a motor 54 to index a desired one into the light path between the distal end of the optical fibers 51 and the photomultiplier tube 52 or similar device. Output signals from the photo detectors 52 are supplied through an A/D converter 55 to a computer 56 provided in the control device 16.
In FIG. 6, it is assumed that the turntable 44 supports, for example thirty cuvettes 45 which are advanced at an interval of, for instance, ten seconds, and the filter unit 53 is rotated by one revolution during these ten seconds. Then, each of the filter elements .lambda..sub.1 to .lambda..sub.10 passes through the light path reaching the photo detector 52 for about one second. During this one second, the drum 49 having formed the slits 48 therein is rotated by one turn. In this manner, absorption data for all wavelengths can be obtained at each photometering position. From these absorption data are selected desired data corresponding to a given wavelength or wavelengths which are determined by the test item, and the selected data are converted into digital values which are then stored in the computer 56. In this manner, for each test liquid in each cuvette on the turntable, the reaction data can be obtained from all photometering positions every ten seconds. Hence, absorbance data for any given test liquid in any given cuvette can be made available at any or all available wavelengths every ten seconds for as long as the cuvette and test liquid remains on the turntable. In the computer, the linear phase when pertinent, can be determined from this data, and thus the kinetic reaction data, if necessary, can be obtained accurately.
As shown in FIG. 9A, the linear phase can be determined at a section near a trigger addition point and having a small value of .vertline.A-B.vertline.. In order to obtain a reaction curve shown in FIG. 9B, it is necessary to compensate for differences in outputs from the photometering positions. To this end, prior to the measurement, a calibrating cuvette having highly accurate optical length is set in the apparatus and absorption values of this cuvette for all wavelengths are measured at all photometering positions and are stored in the computer. During the measurement, the stored absorption values are subtracted from detected values. In this manner, the reaction curve illustrated in FIG. 9B can be obtained.
By increasing the rotation speeds of the rotary filter unit 53 and the slit drum 49, a corresponding increase in data may be obtained from each measuring position.
FIG. 10 is a chart showing the operation of the apparatus according to the invention.
While FIGS. 6 and 7 show two sets of the slit, filter assemblies, and photo detectors, any number of such optical channels may be provided.
It should be noted that since in this embodiment use is made of sequential multi-test mode, it is, of course possible to measure continuously a plurality of test items for each sample as desired by the operator, and supplied to the computer-control device via keyboard, cards or other commonly used computer input devices, etc.
It should be noted that this embodiment offers an operator a number of choices which heretofore would require a sacrifice in productivity and/or convenience to obtain the desired combination of data gathering modes and/or capabilities as follows:
A. Monitoring the change in absorbance of a test liquid over time with capabilities for selectively determining the linear phase of the reaction.
B. Performing such monitoring as in `A.` above at two or more wavelengths.
C. Gathering data for test liquids at only one or two points in time (herein referred to as end-point assays) at one more wavelengths when such desired test liquids are randomly interspersed on the turntable with test liquids requiring data gathering modes as in `A.` or `B.` above.
D. Conversely, test liquids requiring data gathering modes as in `A.` or `B.` above can be randomly interspersed on the turntable with end point assays as in `C.` above.
E. It is further possible to effect continuously a single test item for all samples utilizing any or all of data modes `A.` through `C.` above or;
F. To treat a plurality of samples to a plurality of test items utilizing any or all of the data acquisition modes as in `A.` through `C.` above.
The apparatus of this embodiment further includes the ability for automatic calibration. This can be effected by setting a standard sample to the sample feed mechanism 34 during a stand-by condition. Then, the apparatus automatically operates at every constant time period and the standard sample is delivered into the cuvette 45 on the cuvette feed mechanism 37 and the automatic calibration is effected in a usual manner to compensate for drifts of the apparatus such as variation in brightness of the light source 46, etc.
This automatic calibration ability allows the instrument to be used at any time of day or night with complete confidence that the calibration routine is properly performed regardless of the relative expertise or attention of the operator.
The control of operation of various portions, the inputting operation of patient or sample information, and the calculation of the analyzed results can be effected by a control device (not shown) including one or more computers.
FIG. 11 is a perspective view illustrating an embodiment of the cuvette 45. The cuvette 45 of this embodiment comprises a rectangular opening 45a and a supporting flange 45b provided at the periphery of opening. The opening is connected to a bottom portion 45c by a tapered side wall narrowing towards the bottom portion. The bottom portion 45c is formed as a semicylindrical shape and has measuring windows 45d at both ends, when viewed its axial direction, through which windows the test liquid in the cuvette is optically measured.
According to the above mentioned construction of cuvette 45, since the opening 45a (receiving port) is wide, it is possible to easily deliver the sample and reagent without sputtering them externally. Further, the amount of the test liquid is sufficient to fill the semi-cylindrical bottom portion 45c, and thus the analysis can be effected with very small amounts of the sample and reagent. Moreover, since a measurement axis extends in a longitudinal direction of the cuvette and thus is sufficiently long, it is possible to carry out the analysis with very high sensitivity. Since the side wall is tapered from the opening 45a to the bottom 45c, and the flange 45b is provided around the opening, the cuvette may be simply secured to the cuvette feed mechanism 37 in a manner shown in FIGS. 12A and 12B. That is, the flange 45b may be placed on a holding member 60 as illustrated in FIG. 12A or may be detachably inserted into recesses formed in a holding member 61 as depicted in FIG. 12B. In this manner, the cuvette 45 may be simply supported by the holding member without making the measuring windows 45d in contact with the holding member 60 or 61, and thus the measuring windows can be protected against injury. In FIG. 12A, an arrow E denotes the measuring optical axis. Further, the cuvette 45 may be formed by molding of transparent material, and thus its mechanical strength can be made high.
Next, the sample and reagent delivery mechanisms will be explained. Since these mechanisms can be constructed substantially similarly to each other, only the reagent delivery mechanism will be explained.
As illustrated in FIG. 5 in this embodiment, a plurality of reagent bottles 38 are arranged in the reagent cassette along an endless path. That is to say, as shown in FIGS. 13 and 14, the cassette comprises an elliptic outer frame 80 in which are arranged rotatably a pair of pulleys 81 and 82. An endless belt 83, preferably a timing belt, is arranged between these pulleys 81 and 82. A plurality of partitions 84 are integrally formed with the belt. The reagent bottle 38 is removably inserted into a space formed by adjacent partitions and the outer frame 80. One of the pulleys 81 has formed in its bottom surface, recesses which engage detachably with projection 86 formed on an output shaft of a stepping motor 85 secured to the main body 25. The stepping motor 85 may be driven in either a forward and/or backward direction as determined by means of an externally supplied signal. A handle 87 is secured to stationary shafts of the pulleys 81 and 82 so that the cassette can be easily set into, or taken out of, the holder 32.
In order to maximize the operational efficiency of the analyzing apparatus in which several reagents selected from a number of reagents, are delivered by a single delivery pump, it is preferable to effect the delifery of reagents in such an order that the total traveling distance of the cassette is minimized. For this purpose, the stepping motor 85 for transporting the reagent bottles 38 is of a reversible type.
As illustrated in FIG. 15, information about the order of arrangement of reagent bottles in the cassette has been previously stored in a test item order determining unit. Upon an initiation of measurement for a particular test item, test item data to be effected for the relevant test item is supplied from a memory to the test item order determining unit, to which is also supplied information about a particular reagent bottle which is now in the reagent aspiration position in the reagent bottle transfer device. In the determining unit, the test item order is determined on the basis of these three pieces of information in such a manner that the traveling distance of cassette in the holder can be minimized, and a list for denoting the determined test item order is formed. In accordance with this list the order determining unit controls the successive alignment of reagent bottles with the reagent aspiration station in a sequence so as to insure the optimum economy of movement on the part of the cassette. At the time the list is generated, the list is also supplied to the photometric section, so as to provide the photometric section with test item data relevant to the photometer's responsibilities, for example, the overall sequence of test items and samples on the turntable.
In the above embodiment, all the reagent bottles are arranged in a refrigerator in order to avoid an alteration or deterioration of the reagents. However, some reagents might precipitate under a low temperature and thus should not be stored in the refrigerator. In such a case, as shown in FIG. 16, the cassette holder 32 is divided into two portions, 32A and 32B. One portion (32A) is maintained at room temperature, and reagents which should not be stored at low temperature are installed in this portion 32A. The other portions (32B) is connected to a refrigeration machine 96 and blower 97, to form a closed loop. The operation of the refrigeration machine 96 is controlled by a temperature-detecting element 98 in holder 32B, and a control circuit 99 which receives the output signal from temperature detector 98. It should be noted that the cassette shown in FIGS. 13 and 14 may be installed in portions 32A and 32B. In order to prevent the escape of cool air from refrigerator portion 32B, portion 32B comprises a lid 100 (illustrated in FIG. 17). A small aperture 101 is formed in the lid at a position corresponding to the aspiration position so that a fluid dispensing probe can be inserted into and retracted from portion 32B. The fluid which is used to calibrate the apparatus is preferably stored in refrigerator portion 32B.
In the reagent delivery mechanism as shown in FIG. 18, only a single pump 105 is able to deliver a plurality of different reagents.
In this embodiment, use is made of reagents of high concentration and the reagents are jetted into the cuvettes from the probes together with appropriate diluent(s). By utilizing this construction, the whole apparatus can be made small in size, and contamination between the different reagents can be avoided because the inside of probe is washed by the diluent(s). Since diluent(s) is/are heated to a temperature near the reaction temperature, the temperature of the test liquid can be rapidly increased, and the reaction time can be shortened even if the refrigerated reagent is used and the reaction is carried out in a temperature-controlled incubation environment having a low thermal efficiency, such as an air bath. Further, if the diluent is the same liquid as any required buffer solutions, it is not necessary to provide separate delivery pumps for these liquids.
To begin a dispensing cycle, the desired reagent bottle 38 in the cassette 80 is transported to a position just below the aspiration position of the probe 106. A preheating device 107 is provided to heat the diluent to a temperature near the desired reaction temperature and comprises a heater, a temperature sensor and a temperature control circuit (not shown). The syringe 105 is connected to the probe 106 and a diluent bottle 108 via valves 109 and 110, respectively. In this embodiment, these valves are denoted as two-way valves, but they may be replaced by a single three-way valve. Since these valves 109 and 110 are kept in contact with the diluent only, they do not require chemical resistance. However, in view of a very small amount of liquid to be delivered it is desired that a volume inside the path be kept within very narrow and precise limits. To this end, it is preferable to construct the valves 109 and 110 by use of a rotary solenoid valve with a tapered cock.
The syringe and piston constructing the pump 105 do not require special chemical resistance properties because like the valves 109 and 110 they contact only diluent liquid. In order to deliver different amounts of reagents by the same pump 105 the piston of the pump can be displaced by strokes of variable length by a pulse motor energized by an external signal. As the diluent use may be made of buffer solution as explained above or in some cases use may be made of de-ionized or distilled water.
Operational steps of the reagent delivery pump will be denoted in the following table.
__________________________________________________________________________
Valve 109
Valve 110
Syringe 105
Step Position of probe 106
position
position
piston motion
__________________________________________________________________________
Form air bubble at
Stand by position (in air)
Open Closed
Withdraw slightly
probe tip
Probe into reagent
Stand by position .fwdarw.
Closed
Closed
None
in reagent
Aspirate reagent
In reagent Open Closed
Withdraw to
aspirate reagent
Transport probe
In reagent .fwdarw.
Closed
Closed
None
above cuvette
above cuvette
Deliver reagent and
Above cuvette
Open Closed
Close to dispense
diluent into cuvette
Aspirate diluent
Above cuvette .fwdarw.
Closed
Open Withraw to
stand-by position aspirate diluent
__________________________________________________________________________
When different diluents are used for different reagents, or when a reagent is delivered at several positions, a pluarlity of delivery pumps 105A to 105D may be provided for each diluent as shown in FIG. 19. When a cuvette 45 is transported to a delivery position corresponding to any one of the pumps 105A to 105D, for example 105A, and the reagent to be delivered to this cuvette is that which should be diluted by a diluent connected to this pump 105A, the related reagent bottle 38 is fed to a position corresponding to the pump 105A and then a given amount of the desired reagent is delivered into the cuvette 45 by the pump 105A. On the contrary, if the reagent to be delivered to this cuvette is that which should be diluted by a diluent connected to the pump 105C, after the cuvette is further advanced by two steps, the desired reagent is delivered into the cuvette 45 by the pump 105C.
According to the above explained construction of the reagent delivery mechanism, since the diluents or buffer solutions which are optimum for respective diluents can be used, the reagents can be maintained in a stable condition for a longer time, and the number of possible test items can be increased. For some reagents it is preferable to effect delivery thereof by several stages in order to effect storage of the reagents as component parts in such a way as to prolong the useful chemical stability of these reagents or to dispense quantities which may be beyond the useful dynamic range of a given pump 105A to 105D. In such a case, the same reagent or its component part may be delivered into the same cuvette by a succession of pumps 105A to 105D at successive steps.
In such a discrete delivering operation it is quite important to assure whether a given amount of liquid has been aspirated or not. That is to say, if a serum, sample, or reagent is aspirated excessively or insufficiently, erroneous data would be obtained. Therefore, such a situation must be checked by some means.
FIG. 20A is a schematic perspective view illustrating an embodiment of such means for detecting an amount of aspirated liquid. In this embodiment, the probe 106 is made of